Honeywell SC2000 Stud Sensor User Manual


 
page 74 008-0608-00
10.5.8
SIGNATURE
MODULE
Sub-Menu
UPDATE SIG.MOD?
Menu Item
Updates the information stored in the transducer’s Signature Module with any new
settings that is entered into the instrument with the SETUP menus. The settings that
will be updated are:
EXCITATION
FULL SCALE MV/V
FULL SCALE VALUE
SHUNT CAL VALUE
DISPLAY. UNITS
Note that changes made to the instruments display setup (such as decimal point and
count-by settings) are not updated into the Signature Module.
Updating enters these new settings from the instrument into the Signature Module. If
you do not update the Signature Module, connect a different transducer to the instru-
ment and then re-connect the original transducer again, the instrument will revert to
the settings that are stored inside the Signature Module.
The options for this menu item are:
•“
NO
” will not update the Signature Module.
•“
YES
” will display
WORKING
, update the Signature Module, then display
DONE
.
DEFAULT SIG.MOD?
Menu Item
This menu item will overwrite any user settings in the Signature Module with the origi-
nal default settings. The Signature Module always maintains a copy of the original
default settings that were created at the factory.
The options for this menu item are:
•“
NO
” will not restore original default settings.
•“
YES
” will display
WORKING
, update the Signature Module, with the original factory set-
tings, then display
DONE
.
UPDATE DAC SETUP
Menu Item
This menu item determines whether or not the
DAC. FULL SCALE
and
DAC. ZERO SCALE
Ana-
log Output settings are overwritten with information from a Signature Module when a
new transducer is detected by the instrument.
The options for this menu item are:
•“
YES
” will allow a newly detected Signature Module to update the Analog Output
settings.
•“
NO
” will not allow automatic updating of the Analog Output settings.