Honeywell SC2000 Stud Sensor User Manual


 
page 72 008-0608-00
applied this load to the transducer, press [ENTER].
Upon completion, the display will indicate
DONE
and the instrument will return to the
RUN mode.
If the
CALIBRATION TYPE
is 5-Point Known Load Calibration...
The display will read
DOING 5POINT CAL
, and prompt you to
APPLY 00000 UNIT
(where
00000 and UNIT are previously entered as
KNOWN POINT 1/5
and
DISPLAY. UNITS
).
When you have applied this load to the transducer, press [ENTER].
The display will read
WORKING
, then
APPLY 00250 UNIT
(where 00250 and UNIT are pre-
viously entered as
KNOWN POINT 2/5
and
DISPLAY. UNITS
). When you have applied this
load to the transducer, press [ENTER].
The display will read
WORKING
, then
APPLY 00500 UNIT
(where 00500 and UNIT are pre-
viously entered as
KNOWN POINT 3/5
and
DISPLAY. UNITS
). When you have applied this
load to the transducer, press [ENTER].
The display will read
WORKING
, then
APPLY 00750 UNIT
(where 00750 and UNIT are pre-
viously entered as
KNOWN POINT 4/5
and
DISPLAY. UNITS
). When you have applied this
load to the transducer, press [ENTER].
The display will read
WORKING
, then
APPLY 01000 UNIT
(where 01000 and UNIT are pre-
viously entered as
KNOWN POINT 5/5
and
DISPLAY. UNITS
). When you have applied this
load to the transducer, press [ENTER].
The display will now read
RETURN PNTS.=NO
. If you press [ENTER], the display will
indicate
DONE
and the instrument will return to the RUN mode. If you use the [UP]
button and [ENTER] to select
RETURN POINTS=YES
, you may now apply the same
loads again but in reverse order. This will allow the instrument to measure and
partially eliminate any hysteresis that may be present in the transducer. After the
loads have been applied again, the display will indicate
DONE
and the instrument
will return to the RUN mode.