Performance Tests
TDS 310, TDS 320 & TDS 350 Service Manual
4Ć15
13. Check that the PkĆPk readout on screen (as shown in
FigureĂ4Ć6) is w42.5 mV.
Second, read the results
from the readout of
measurement PkĆPk.
First, increase the reference
frequency to the test
frequency; then decreasethe
horizontal scale.
Figure 4Ć6:ăMeasuring Analog Bandwidth
14. When finished checking, set the SEC/DIV back to the 10 ms setting, and
set the generator output frequency back to 50 kHz.
15. Press WAVEFORM OFF to remove Channel 1 from the display.
16. Press CH 2 and move the hookup to the CH 2 input.
17. Press TRIGGER MENU ! Source ! CH 2.
18. Repeat steps 6 through 13 for CH 2.
19. Disconnect the test hook up from the CH 2 input connector.