Specifications
Agilent 81130A Specifications
Deskew
Compensation for different cable delays.
For frequencies >170 kHz only.
Transition Times
Measured between 10% and 90% of amplitude, except for ECL levels (20%
and 80% of amplitude).
Digital Channel Add
In this mode, channel 1 and channel 2 are added and fed to channel 1
output. Channel 2 is still available.
Deskew
Agilent 81130A with
Agilent 81131A
Agilent 81130A with
Agilent 81132A
Range:
±
25 ns
Resolution: 4 digits, 2 ps best case
Transition Times
Agilent 81130A with
Agilent 81131A
Agilent 81130A with
Agilent 81132A
Range:
800 ps or 1600 ps
(selectable)
fixed
Minimum transition:
600 ps for Vpp
1 V
900 ps for Vpp > 1 V
500 ps typ.
At ECL levels: <450 ps < 350 ps (200 ps typ.)