Agilent Technologies Agilent 81130A 400 Welding System User Manual


 
63
Programming Reference
SCPI Instrument Command List
Command
:CURR[1|2]:LOW
Long
[:SOURce]:CURRent[1|2][:LEVel][:IMMediate]:LOW
Form
Set & Query
Parameter
Numeric
Parameter suffix
A
with engineering prefixes.
*RST value
10 mA (50 into 50 )
Specified Limits
3.8V Outputs (50 into short): max. 152 mA typical
3.0V Outputs (50 into short): max. 120 mA typical
Value coupling
Range coupling
High-level
Description
This command programs the Low-level current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute the
[:SOURce]:HOLD CURRent
command to enable the
[:SOURce]:CURRent
subsystem.
The available current range is limited by the specified voltage limits.
Example
:HOLD CURR
Enable CURRENT subsystem
:CURR1:LOW 50 MA
Set OUTPUT 1 Low-level to 50 mA
Command
:CURR[1|2]:LIM
Long
[:SOURce]:CURRent[1|2]:LIMit[:HIGH]
Form
Set & Query
Parameter
Numeric
Parameter suffix
A
with engineering prefixes.
*RST value
+10.0 mA
Amplitude = High – Low
Offset =
High – Low
2