Agilent Technologies Agilent 81130A 400 Welding System User Manual


 
39
Programming Reference
Programming the Instrument Trigger Modes
Continuous
Set Continuous mode by arming the instrument from its internal PLL:
:ARM:SOURce IMMediate Arm from internal PLL
Started
Set Started mode by arming the instrument on low to high level transition
from the EXT INPUT:
:ARM:SOURce EXTernal1 Arm from EXT INPUT
:ARM:MODE STARted Start on the arm event
:ARM:SENSe POSitive Arm on positive (high) level
:ARM:LEVel:THReshold 1V Set EXT INPUT threshold
Gated
Set Gated mode by arming the instrument on levels from the EXT INPUT:
:ARM:SOURce EXTernal1 Arm from EXT INPUT
:ARM:MODE GATed Select gated mode
:ARM:SENSe POSitive Arm on positive level
Pulses
Set Pulses mode by setting the
:TRIGger:COUNt
to 1 so that a single
triggered pulse period is generated for every :ARM event. The trigger
source sets the pulse period:
:TRIGger:COUNt 1 Single pulse period per arm event
:TRIGger:SOURce INTernal 1 Pulse period from internal PLL
:DIGital:PATTern OFF Disable pattern data.
Pulse period source :TRIGger SOURce
internal PLL
CLK-IN
INTernal[1]
or
IMMediate
EXTernal2