Agilent Technologies Agilent 81130A 400 Welding System User Manual


 
43
Programming Reference
SCPI Instrument Command List
Command
:ARM:LEV[:THR]
Long
:ARM[:SEQuence[1] | :STARt][:LAYer]:LEVel[:THReshold]
Form
Set & Query
Parameter
Numeric
Parameter Suffix
V
with engineering prefixes.
*RST value
+1.0 V
Specified Limits
1.4 V to +3.7 V
Description
Use this command to program the triggering threshold of the EXT INPUT
connector.
Example
:ARM:LEV 2.5V
Set EXT INPUT threshold to 2.5 V
Command
:ARM:LEV:TERM
Long
:ARM[:SEQuence[1] | :STARt][:LAYer]:LEVel:TERMination
Form
Set & Query
Parameter
Numeric
Parameter Suffix
V
with engineering prefixes.
*RST value
+0.0 V
Specified Limits
2.1 V to +3.3 V
Description
Use this command to program the termination voltage compensation of
the EXT INPUT connector.
Example
:ARM:LEV:TERM 1.0V
Set EXT INPUT termination voltage to 1.0 V
Command
:ARM:MODE
Long
:ARM[:SEQuence[1] | :STARt][:LAYer]:MODE
Form
Set & Query
Parameter
STARted | GATed
*RST value
STARted