54
Programming Reference
SCPI Instrument Command List
Description
Use this command to set the type of the segment for one channel.
If the segment type of one channel is set to PRBS the other channel may
not be set to DATA.
If at least one channel uses PRBS, then the segment type combination
used in this segment has to be used in every segment that shall generate a
PRBS.
Command
:DIG:PATT:UPD
Long
:DIGital[:STIMulus]:PATTern:UPDate
Form
Set & query
Parameter
ON | OFF | ONCE
*RST value
ON
Description
Use this command to enable and disable the automatic updating of the
pattern generating hardware following a
:DIG:PATT:SEGM[1|2|3|4]:DATA
command. Disable the automatic
updating if you want to set up new pattern data in the instrument without
affecting the pattern which is currently being generated. You can then
update the hardware with the new pattern data by sending a
:DIG:PATT:UPD ONCE
command.
Command
:DIG:SIGN[1|2]:FORM
Long
:DIGital[:STIMulus]:SIGNal[1|2]:FORMat
Format
Set & Query
Parameter
RZ | NRZ | R1
Range Coupling
Period, Frequency
*RST value
RZ