Agilent Technologies Agilent 81130A 400 Welding System User Manual


 
62
Programming Reference
SCPI Instrument Command List
Command
:CURR[1|2]:HIGH
Long
[:SOURce]:CURRent[1|2][:LEVel][:IMMediate]:HIGH
Form
Set & Query
Parameter
Numeric
Parameter suffix
A
with engineering prefixes.
*RST value
+10 mA (50 into 50 )
Specified Limits
3.8 V Outputs (50 into short): max. 152 mA typical
3.0 V Outputs (50 into short): max. 120 mA typical
Value coupling
Range coupling
Low-level
Description
This command programs the High-level current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute
[:SOURCE]:HOLD CURRent
command to enable the
[:SOURCE]:CURRent
subsystem.
The available current range is limited by the specified voltage limits.
Example
:HOLD CURR
Enable CURRENT subsystem
:CURR1:HIGH 150MA
Set OUTPUT 1 High-level to 150 mA
Amplitude = High – Low
Offset =
High – Low
2