Agilent Technologies Agilent 81130A 400 Welding System User Manual


 
58
Programming Reference
SCPI Instrument Command List
Command
:MMEM:INIT
Long
:MMEMory:INITialize
Form
Event
Parameter
["A:"[,"DOS"]]
*RST value
Not applicable
Description
Use this command to initialize a memory card to DOS format.
CAUTION
Initializing a memory card destroys any existing data on the card.
Command
:MMEM:LOAD:STAT
Long
:MMEMory:LOAD:STATe
Form
Event
Parameter
<n>,"filename"[,"A:"]
*RST value
Not applicable
Specified Limits
<n> = 0 to 4 (integer)
Description
Use this command to load a complete instrument setting from file
filename in the current directory into memory <n> in the instrument.
Memories 1 to 4 are the internal memories. Use memory 0 to load a
setting as the current instrument setting.
Examples
See next command
Command
:MMEM:STOR:STAT
Long
:MMEMory:STORe:STATe
Form
Event
Parameter
<n>,"filename"[,"A:"]
*RST value
Not applicable