Agilent Technologies Agilent 81130A 400 Welding System User Manual


 
50
Programming Reference
SCPI Instrument Command List
Description
Use this command to set or read a segments data of one or all channels
starting from Bit 1. The <data> is an arbitrary block of program data as
defined in IEEE 488.2 7.7.6.2, for example:
#1511213
#
Start of block
1
Length of the length of the data
5
Length of the data
11213
5 bytes of data
#2161000100010001000
#
Start of block
2
Length of the length of the data
16
Length of the data
10...00
16 bytes of data
#011213
#
Start of block
0
Replaces the data block length specification. Length is
calculated automatically.
11213
5 bytes of data
NOTE
The data length meets the same restrictions, than the segment length
(see page 52).
Example
:DIG:PATT:SEGM1:DATA #1511213