Agilent Technologies Agilent 81130A 400 Welding System User Manual


 
45
Programming Reference
SCPI Instrument Command List
Command
:INIT:CONT
Long
:INITiate:CONTinuous
Form
Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
ON
Description
Use this command to enable/disable automatic restart of the instrument
(equal to start and stop the instrument). If
:ARM:SOURce
is set to
IMMediate
, the value of
:INITiate:CONTinuous
is ignored.
Command
:CHAN:MATH
Long
:CHANnel:MATH
Form
Set & Query
Parameter
OFF | DIGital
*RST value
OFF
Description
Use this command to enable or disable digital channel addition in an
instrument with two Output channels installed.
With
:CHAN:MATH DIGital
the digital signals from both channels are
xored (before the slopes are applied) at OUTPUT 1. The signal of
OUTPUT 2 can be used in parallel.
This allows you to for example to simulate single or repeated glitches.
Command
:DIG:PATT:LOOP
Long
:DIGital[:STIMulus]:PATTern:LOOP[:LEVel[1]][:COUNt]
Form
Set & Query
Parameter
Numeric
*RST value
1
Specified Limits
1 to 2^20