Hansol Multitech Uromat 3000 Welding System User Manual


 
06/00 - 11 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
1.8 Power Line Connection Data
Power lead-in must be routed over a 30 mA fault current interrupter that is provided by the
customer. The room installation must comply with VDE 0107.
In all other countries outside the Federal Republic Germany, the legally specified country
regulations take precedence and must be maintained.
Prerequisite:
The unit is designed for single-phase DC current with a fixed installation and for a fixed
connection using an all-pole separator from the Network (IEC 601, Chapter 57.1).
During installation, it must be possible to adapt the power line voltage and frequency to
correspond to the order.
Power connection: 1 N 115 /200/208/ 230/240 V AC
Frequency: 50 / 60 Hz
Nominal current (fuse): 13/7.5/7.2/6.5/6.25 A
Nominal line power: 1500 VA
Heat dissipation: 240 W
1.9 Required Special Tools
Torque wrench 50 Nm (5 mkp)
Masonry drill bit 12 mm dia.
Sista sanitary caulking F 101
Special grease - Tunap Tungrease BS Pa. Nr.: 9026 0001
1.10 Required Test Equipment
PC or laptop with color display (black/white is also sufficient, but operation is made more difficult);
min. 486 processor; 16 MB RAM; with a 3 ½" diskette drive and hard drive.
Windows 3.x ; Win 95 operating system (Windows NT can cause difficulties with the Com interface).
Serial connection cable (9-pole, Sub-D connector to 9-pole Sub-D socket, 1-1 connection), length > 1m.
Spirit level
Tape measure
Multimeter test instrument
1.11 Conditions for Transport and Storage
Ambient temperature range -25 C to 70 C
Relative humidity in the range 5% to 95%
Air pressure in the range 700 hPa to 1100 hPa
1.12 Operating Conditions
Ambient temperature range 10 C to 40 C
Relative humidity in the range 20% to 80%
Air pressure in the range 700 hPa to 1100 hPa