Hansol Multitech Uromat 3000 Welding System User Manual


 
06/00 - 85 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
F 40 Format switch cassette incorrect. Check 4 format switches.
F 41 M1 motor breaker stuck/defective. Check K1 breaker.
F 42 M2 motor breaker stuck /defective. Check K2 breaker.
F 43 M3 motor breaker stuck /defective. Check K3 breaker.
F 44 M4 motor breaker stuck /defective. Check K4 breaker.
F 45 M5 motor breaker stuck /defective. Check K5 breaker.
F 46 M6 motor relay stuck /defective. Check K19 relay (CPU board).
F 47 FU0 N>0 active. FU0 relay defective/ input
defective / check cabling.
F 48 FU1 N>0 active. FU1 relay defective / input
defective / check cabling.
F 50 Collimator malfunction. Check cabling, potentiometers,
setup parameters.
F 61 Limit switch, temperature, breaker
malfunction, M1 tilt /vertical drive or M3.
Free run (press button on breaker
board). Allow motor to cool down.
F 62 Limit switch, temperature, breaker
malfunction, M2 system drive.
Free run (press button on breaker
board). Allow motor to cool down.
F 64 Limit switch, temperature, breaker
malfunction, M4 table longitudinal drive.
Free run (press button on breaker
board). Allow motor to cool down.
F 79 Invalid parameter version Initialize correct parameter
version.
F 80 EEPROM memory error. Download new parameter file.
F 99 Unit is in the service mode. Set the service switch to the RUN
mode.
F = Fatal Error = System error = cannot be corrected