Hansol Multitech Uromat 3000 Welding System User Manual


 
06/00 - 61 of 90 - 0116 7201
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
f. Display Self Test
After switching on power, all segments in the tilt display and the large round LED’s are
tested individually. If this does not happen, there is a malfunction.
Seven-segment display: visible by a circular sequence of the active segments. All segments
must light up. If a segment does not go on, there is a malfunction.
LED’s: All three LED’s light up cyclically in the three colors: red, green and yellow. If one
color does not appear, there is a malfunction.
g. Footswitch Self Test
This test cannot be viewed. However, correct function is signaled via the bus.
h. Control Unit Self Test
After switching on power, all LED’s are individually tested by blinking, the multi-colored
LED’s also blink in the three colors. If this does not happen, there is a malfunction.
3.24 Startup
Adjusting the Axes
If a PC is not available, proceed as described below:
Switch the unit with the switch (Fig. 76/Pos. 1) into the service mode, the green LED (Pos. 2)
on the controller goes on and stays on.
Press the Store and Reset buttons on the control unit at the same time, the three memory
LED`s on the control unit blink green and the LED on the controller blinks very quickly.
Each axis can now be moved normally. Exception: the cassette can be moved slowly forward
and backward by pressing the Recall and Reset buttons.
To save a new software limit, Store must be pressed and held (M1 LED blinks red) and the
corresponding direction button for the desired movement must also be pressed.
If the controller has recognized this correctly, the M1 LED blinks red, and the position has
been accepted and the keys can be released again. The M1 LED then blinks green again.
To save a centered position, the Store button must again be pressed and held and then the
two buttons for the corresponding axis must either be pressed and held, or pressed in
sequence.
To move past an incorrectly set software limit, the ZOOM button must be held down and the
corresponding direction button must also be held. The software limit is now set to the
maximum AD value.
If an axis is completely adjusted, the new limits should be saved in the EEprom. To do this,
the Store and Reset buttons must be pressed at the same time. During the save routine
(approx. 5 sec) all three memory LED’s will blink red.
When the save routine is completed, the memory LED`s will again blink green.