Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
3.27 Saving the Parameters in the PC
Switch off the unit. Connect the PC and the unit using the connection cable included in the
shipment.
Switch the unit into the service mode with the switch (Fig. 77/Pos. 1), the green LED (Pos. 2)
controller goes on continuously.
Start the URO service tool. See 3.23
Open the parameter file included in the shipment and read the parameters from the Uromat
and save them (do not overwrite the original file).