1. TECHNICAL DATA
1.1 General Safety Information
Maintenance and repair may be performed only by an office that is authorized by the
manufacturer.
In the Federal Republic of Germany, electrical installation of medically used rooms must
conform to VDE Regulation 0107. In all other countries, the particular applicable national
regulations take precedence and must be observed. These can be found in the system
project plan.
During installation, it must be assured that all ground wire connections provided by the
manufacturer must be connected prior to starting up the unit.
The ground wire between the individual components and the power supply must be
connected as shown in the Wiring Diagram.
The regulations provided by the trade unions for occupational safety and prevention of
accidents must be observed.
No work may be performed on parts that are under power (above 42 V).
This condition applies for both measurement and adjustment work steps. However, special
care is required when performing these worksteps.
If the operating instructions call for voltage to be switched on for unit movements, the system
must be shut down immediately following such tests.
The radiation regulations must be observed while making settings and checks that must be
performed under X-radiation; radiation protective clothing must be worn.
1.2 Component Weights
Unit column approx. 320 kg
Table approx. 230 kg
Tube unit support arm approx. 50 kg
Footswitch approx. 5 kg
Total weight, incl. tube unit, collimator,
adjustable Bucky and I.I.: approx. 730 kg
1.3 Component Dimensions
Unit column with table and tube unit support arm on the pallet.
2050 mm x 1015 x 1800 mm
1.4 Packaging Dimensions and Weights
Unit column with table, incl. accessories 2050 mm x 1015 mm x 1890 mm
approx. 900 kg
Transport carriage 1800 mm x 200 mm x 200 mm
approx. 85 kg