2.6 Temporarily Connecting Line Power
Caution:
If the unit is connected to a different line power than that listed on the order, the setting must
be changed to the one that corresponds to the local line power as specified by the Wiring
Diagram to the power line transformer (Fig. 12/Pos. 1).
Provide the unit with power by connecting a temporary connection cable (plug-in cable) to
move the table for individual installation steps.
Connect the multi-function footswitch on the bottom of the table.
Fig. 13 Fig. 14
2.7 Installing the Tube Unit Support Arm
Tilt the tabletop all the way down into the horizontal position and move it towards the front,
transversely to the user, until it is in the end position. Move the tube unit carriage (system
carriage) max. to the head end.
Remove the drain panel (Fig. 13/Pos. 1) after loosening the 5 mounting screws (Pos. 2,3+4).
Place the tube unit support arm (Fig. 14/Pos. 1) on the longitudinal carriage (Pos. 2) and
secure it in place with the 6 screws (Pos. 3+4).
Remove the red transport bracket (Fig. 05/Pos. 1) (hooking device).
Reinstall the drain panel (Fig. 13/Pos. 1).