Hansol Multitech Uromat 3000 Welding System User Manual


 
0116 7201 -84 of 90 - 06/00
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
5. Troubleshooting
Error
Number
Cause
Corrective Measure
F 1 Frequency converter 0 for tilt/vertical,
system failed.
Check connection cables, settings
(address, Baud rate).
F 2 Frequency converter 1 for tilt/vertical, table
longitudinal movement, cassette failed.
Check connection cables, settings
(address, Baud rate).
F 3 Frequency converters 0+1 failed. Check connection cables, settings
(address, Baud rate).
F 10 Footswitch communication problem. Check fuse, cabling.
F 11 Manual control unit communication
problem.
Check fuse, cabling.
F 13 Incorrect footswitch software version. Replace footswitch.
F 14 Incorrect manual control unit software
version.
Replace manual control unit.
F 20 Analog voltage failed. Check fuse.
F 21 M1 tilt/vertical potentiometer error. Check potentiometer cables and
potentiometer.
F 22 System potentiometer error. Check potentiometer cables and
potentiometer.
F 23 M3 tilt/vertical potentiometer error. Check potentiometer cables and
potentiometer.
F 24 Table longitudinal potentiometer error. Check potentiometer cables and
potentiometer.
F 25 Cassette potentiometer error. Check potentiometer cables and
potentiometer.
F26 Table transverse potentiometer error /
motor defective.
Check motor potentiometer cables
and potentiometer.
F 28 Collimator vertical potentiometer error. Check potentiometer cables and
potentiometer.
F 30 Button fuse malfunction. Check buttons, cables.
F 31 Footswitch button sticking. Check buttons.
F 32 Manual control unit button sticking. Check buttons.
F 33 Exposure switch active during system start. Check switches.
F 34 Fluoroscopy switch active during system
start.
Check switches.
F 35 Exposure centering switch defective. Check switches, cabling.