0116 7201 -40 of 90 - 06/00
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
2.8 Preinstalling the Tube Unit – Collimator Combination
Assemble the tube unit – Collimator combination according to the description in the
manufacturer's installation instructions. See Fig. 15
Fig. 15 Fig. 16
2.9 Installing the Tube Unit – Collimator on the Tube Unit Support Arm
Tilt the tabletop until it is vertical and move the tube unit carriage towards the foot-end.
Remove the cover (Fig. 18/Pos. 6) on the tube unit support arm. Remove the support studs
(Fig. 16/Pos. 1) from the tube unit support arm (Pos. 2). Place the preinstalled tube unit –
collimator combination (Fig. 17/Pos. 1) down on its side and move the command arm (Pos.
2) between the support studs (Pos. 3) and the tube unit housing (Pos. 1) and secure them in
place with the 4 screws (Pos. 5).
Fig. 17 Fig. 18