Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf-Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru
Place the assembled tube unit – collimator combination (Fig. 17/Pos. 1) on the tube unit
support arm (Fig. 18/Pos. 2) using the support studs (Pos. 3), insert the shims (Fig. 16/ Pos.
5) and secure it in place using the 4 screws (Pos. 6).
2.10 Installing the Image Intensifier
Remove the I.I. adapter plate (Fig. 20/Pos. 1) from the I.I. carriage (Pos. 2). Secure the I.I.
adapter plate (Fig. 19/Pos. 1) in position on the I.I. housing (Pos. 2) with the screws. Install
the I.I. unit on the I.I. carriage (Fig. 20/Pos. 2) using the two strips (Pos. 4+5) the 8 washers
and the 8 nuts (Pos. 7). Max. torque: 20 Nm.
Fig. 19 Fig. 20
2.11 Laying the Collimator Cables
The cable for the collimator is laid in the unit base parallel to the high voltage cables. The
cable for the collimator is laid parallel from the unit controller in the unit base to the high
voltage cables.
Clamp, or plug in, the cable (Fig. 12/Pos. 2) on the unit controller (Pos. 3),at the connection
points provided.
Lay the cables later in Section 2.12, Section 2.12.