NARDA 8718B Stud Sensor User Manual


 
26 Operating the Meter
To use a correction factor for a probe NOT in memory...
1. Press F3 (MENU). The First Menu Screen will appear.
2. Press 8 (NEXT) to access the Second Menu Screen.
3. Press 1 (COR FACT). A screen will appear that shows
the correction factor.
4. Use the correction factors on the handle of the probe to
determine the appropriate correction factor and then
use the numeric keys on the meter to enter this value.
Enter a three digit number, ignoring the decimal point,
and rounding to two digits to the right of the decimal
point. If you make a mistake, just enter more digits
until correct. For example, to enter a correction factor
of 1.22, simply enter 1, 2, and then 2 while a correction
factor of 0.94 is entered as 0, 9, and 4.
5. Press Enter to save this correction factor. The meter
automatically returns to the measurement screen.
6. The numeric correction factor will be displayed on
the right side of the third line of the display in the form
CF X.XX.
1 DATA LOG 5 UNI TS
2 TIME AVG 6 RS232
3 SPAT I AL 7 PROBE
4 BAT/LIFE 8 NEXT
F1 F2 F3 F4
1 COR FACT 5 ALARM
2 SETT I NGS 6 STNDS
3 CABLE/ FO 7 BLANK
4 LOCKOUT 8 BACK
F1 F2 F3 F4