NARDA 8718B Stud Sensor User Manual


 
58 Operating the Meter
Mode 1 Averaging
To initiate Mode 1 averaging (fixed time averaging over any
one of several specific time intervals)...
1. Press number 1 on the keypad to select FIXED TIME
from the averaging menu screen. Selecting Mode 1
averaging will automatically shut off Mode 2 averaging
if it is on.
2. A new menu screen will appear that offers four
different averaging intervals: 30 SEC, 1MIN, 6MIN,
and 30 MIN. If an averaging interval had previously
been selected and is currently active, that selection will
be flashing.
3. Press the number on the numeric keypad (1-4) that
corresponds to the desired averaging interval and the
meter automatically returns to the measurement
screen with new options available for the function
keys. If you had accessed the menu system from the
Opening Menu Screen, you will return to that screen.
4. At any time you can press F4 (QUIT) to turn the time
averaging mode off and return to the normal
measurements screen.
F I XED T IME AVERAGING
1=30 SEC 3= 6 MIN
2= 1 MIN 4=30 MIN
OFF
F1 F2 F3 F4
12. 1 % STD
27.5 MAX CF 1.00
START MENU QU I T
F1 F2 F3 F4