120
On Page 17, in Voltage Monitor and Remote Readback Full Scale Calibration step d, change the A8R58 adjustment to:
170.025
±6mV (6032A)
50.0075
± 1.52mV (6032A)
425.063
± 6µV (6035A)
On Page 17, in Voltage Monitor and Remote Readback Full Scale Calibration step c, change 5.000625V to:
4.250625V
± 100µV (6030A/35A)
4.167187V
± 100µV (6032A)
On Page 17, in Voltage Monitor and Remote Readback Full Scale Calibration step f, change the A8R61 adjustment to
between:
170.00 and 170.50Vdc (6030A)
50.00 and 50.025Vdc (6032A)
425.00 and 425.125Vdc (6035A)
On Page 20, in Power Limit Calibration step a, change 240 Vdc to 206 Vdc.
On Page 20, in Power Limit Calibration step b, change the load resistor as follows:
2.9 ohm 1500 W (6030A)
0.05 ohm 1500 W (6031A)
0.29 ohm 1500 W (6032A)
32 ohm 1500 W (6035A)
On Page 20, in Power Limit Calibration step e, change the VSET values to:
VSET 50 (6030A)
VSET 6 (6031A)
VSET 14.5 (6032A)
VSET 163 (6035A)
On Page 20, in Power Limit Calibration step f, change the output values to:
50
± 1V (6030A)
6
± 0.08V (6031A)
14.5
± 0.1V (6032A)
163
± 1V (6035A)
On Page 20, in Power Limit Calibration step g, change the current and resistor settings to:
32 ohm 1500W (6030A)
0.50 ohm 1500W (6031A)
2.8 ohm 1500W (6032A)
212 ohm 1500W (6035A)
On Page 20, in Power Limit Calibration step i, change the VSET or ISET values to:
VSET 170 (6030A)
ISET 40 (6031A)
VSET 50; ISET 18 (6032A)
VSET 425 (6035A)
On Page 21, in Power Limit Calibration step j, change the output values to:
170
± 2V (6030A)
@40A (6031A)
50
± 0.4V @17.5A (6032A)
425
± 0.4V @2A (6035A)