Agilent Technologies 6030A Welding System User Manual


 
122
On Pages 25-27, in RMS Measurement Procedure, Peak Measurement and Load Transient Recovery Time, change the
output voltage to
42Vdc (6030A)
5.2Vdc (6031A)
13.5Vdc (6032A)
150Vdc (6035A)
On Page 28 and 29, in paragraph Current Programming and Readback Accuracy steps c and g, change the VSET values to:
VSET 170 (6030A)
VSET 50 (6032A)
VSET 425 (6035A)
On Page 29, in Load Effect (Load Regulation) step c, change the output current to:
4.7Adc (6030A)
35Adc (6032A)
16Adc (6035A)
1.38Adc (6035A)
On Page 29, in Load Effect (Load Regulation) step cd change the output voltage to:
170Vdc (6030A)
50Vdc (6032A)
425Vdc (6035A)
On Page 30, in Source Effect (Line Regulation) step d, change the output current to:
17Adc (6030A)
120Adc (6032A)
50Adc (6035A)
5Adc (6035A)
On Page 30, in Source Effect (Line Regulation) step e, change the output voltage to:
42Vdc (6030A)
5.2Vdc (6031A)
13.5Vdc (6032A)
150Vdc (6035A)
On Page 30, in PARD Ripple and Noise, make the following resistor changes:
from 3.5 to 2.4 ohms (6030A)
from 0.069 to 0.043 ohms (6031A)
from 0.4 to 0.27 ohms (6032A)
from 40 to 30 ohms (6035A)
On Page 30, Initialization, change step 2 as follows:
2. Turn the power on, and depending on your unit's model number, send the following strings:
"EEINIT 6030" ''EEINIT 6031''
"EEPROM 10,4,175.00" ''EEPROM 10,4,17.5"
''EEINIT 6032" "EEINIT 6035"
"EEPROM 10,4,51.495" "EEPROM 10,4,437.50"