Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 213
:MARKer:YDELta
(see page 530)
Query Syntax
:MARKer:YDELta?
The :MARKer:YDELta? query returns the value difference between the
current Y1 and Y2 cursor positions.
Ydelta = (Value at Y2 cursor) - (Value at Y1 cursor)
Return Format
<value><NL>
<value> ::= difference value in NR3 format
See Also • "Introduction to :MARKer Commands" on page 204
• ":MARKer:MODE" on page 205
• ":MARKer:X1Y1source" on page 207
• ":MARKer:X2Y2source" on page 209
• ":MARKer:Y1Position" on page 211
• ":MARKer:Y2Position" on page 212
NOTE
If the front-panel cursors are off or are set to Binary or Hex Mode, the marker position
values are not defined. Make sure to set :MARKer:MODE to MANual to put the cursors in
the front-panel Normal mode.