Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 343
:TRIGger[:EDGE]:LEVel
(see page 530)
Command Syntax
:TRIGger[:EDGE]:LEVel <level>
<level> ::= <level>[,<source>]
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= 2 V with probe attenuation at 1:1 in NR3 format for
external triggers
<source> ::= {CHANnel<n> | EXTernal}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger[:EDGE]:LEVel command sets the trigger level voltage for the
active trigger source.
Query Syntax
:TRIGger[:EDGE]:LEVel? [<source>]
The :TRIGger[:EDGE]:LEVel? query returns the trigger level of the current
trigger source.
Return Format
<level><NL>
See Also • "Introduction to :TRIGger Commands" on page 313
• ":TRIGger[:EDGE]:SOURce" on page 346
NOTE
If the optional source is specified and is not the active source, the level on the active source
is not affected and the active source is not changed.