334 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3 Commands by Subsystem
CAN Id specification is set by the :TRIGger:CAN:PATTern:ID
and:TRIGger:CAN:PATTern:ID:MODE commands.
CAN Data specification is set by the :TRIGger:CAN:PATTern:DATA
command.
CAN Data Length Code is set by the :TRIGger:CAN:PATTern:DATA:LENGth
command.
Query Syntax
:TRIGger:CAN:TRIGger?
The :TRIGger:CAN:TRIGger? query returns the current CAN trigger on
condition.
Return Format
<condition><NL>
<condition> ::= {SOF | DATA | ERR | IDD | IDE | IDR | ALL | OVER | ACK}
Errors "- 241, Hardware missing" on page 501
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:MODE" on page 319
":TRIGger:CAN:PATTern:DATA" on page 326
":TRIGger:CAN:PATTern:DATA:LENGth" on page 327
":TRIGger:CAN:PATTern:ID" on page 328
":TRIGger:CAN:PATTern:ID:MODE" on page 329
":TRIGger:CAN:SIGNal:DEFinition" on page 496
":TRIGger:CAN:SOURce" on page 332
NOTE
SOF is the only valid selection for analog oscilloscopes. If the automotive CAN and LIN
serial decode option (Option AMS) has not been licensed, SOF is the only valid selection.