:MEASure:VAVerage
[<source>] (see
page 247)
:MEASure:VAVerage?
[<source>] (see
page 247)
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= calculated
average voltage in NR3 format
:MEASure:VBASe
[<source>] (see
page 248)
:MEASure:VBASe?
[<source>] (see
page 248)
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<base_voltage> ::= voltage at the
base of the selected waveform in
NR3 format
:MEASure:VMAX
[<source>] (see
page 249)
:MEASure:VMAX?
[<source>] (see
page 249)
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= maximum
voltage of the selected waveform
in NR3 format
:MEASure:VMIN
[<source>] (see
page 250)
:MEASure:VMIN?
[<source>] (see
page 250)
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= minimum
voltage of the selected waveform
in NR3 format
:MEASure:VPP
[<source>] (see
page 251)
:MEASure:VPP?
[<source>] (see
page 251)
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= voltage
peak-to-peak of the selected
waveform in NR3 format
:MEASure:VRMS
[<source>] (see
page 252)
:MEASure:VRMS?
[<source>] (see
page 252)
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= calculated dc
RMS voltage in NR3 format
n/a :MEASure:VTIMe?
<vtime>[,<source>]
(see page 253)
<vtime> ::= displayed time from
trigger in seconds in NR3 format
<return_value> ::= voltage at the
specified time in NR3 format
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
Table 12 :MEASure Commands Summary (continued)
Command Query Options and Query Returns