CALibrate:PASSword
This command can only be used in calibration mode. It allows you to change the calibration password. A
new password is automatically stored in nonvolatile memory and does not have to be stored with
CALibrate:SAVE. If the password is set to 0, password protection is removed and the ability to enter the
calibration mode is unrestricted.
Command Syntax
CALibrate:PASSword <NRf>
Parameters
0 (default)
Examples
CAL:PASS 6812 CAL:PASS 02.1997
Related Commands
CAL:STAT
CALibrate:PWM:FREQuency
Agilent 6811B, 6812B, 6813B Only
This command is only used during manufacture or repair. It trims the switching frequency of the power
output stages. The numbers from 0 to 7 are internally mapped to 8 discrete frequencies.
Command Syntax
CALCulate:PWM:FREQuency <NRf>
Parameters
0 through 7
Examples
CAL:PWM:FREQ 1
Query Syntax
CALibrate:PWM:FREQuency?
Returned Parameters
<NR1>
Related Commands
CAL:PWM:RAMP
CALibrate:PWM:RAMP
Agilent 6811B, 6812B, 6813B, Only
This command modulates the slope of voltage ramp driving the power output stages. Varying the ramp
affects the harmonic distortion of the output. The argument is a number from 0 to 255. This command is
only used during manufacture or repair of the .
Command Syntax
CALCulate:PWM:RAMP <NRf>
Parameters
0 through 255
Examples
CAL:PWM:RAMP 100
Query Syntax
CALibrate:PWM:RAMP?
Returned Parameters
<NR1>
Related Commands
CAL:PWM:FREQ
CALibrate:SAVE
This command can only be used in calibration mode. It saves any new calibration constants (after a
current or voltage calibration procedure has been completed) in nonvolatile memory.
Command Syntax
CALibrate:SAVE
Parameters
None
Examples
CAL:SAVE
Related Commands
CAL:CURR CAL:VOLT CAL:STAT