Agilent Technologies 6814B Welding System User Manual


 
Language Dictionary - 3
95
Bit Configuration of Operation Status Registers
Bit Position 15–9 8 7–6 5 4–1 0
Bit Name not used CV not used WTG not used CAL
Bit Weight 256 32 1
CAL = Interface is computing new calibration constants
WTG = Interface is waiting for a trigger.
CV = Output voltage is regulated.
STATus:OPERation?
This query returns the value of the Operation Event register. The Event register is a read-only register
which holds (latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the
Operation Event register clears it.
Query Syntax
STATus:OPERation[:EVENt]?
Parameters
None
Examples
STAT:OPER:EVEN?
Returned Parameters
<NR1> (register value)
Related Commands
*CLS STAT:OPER:NTR STAT:OPER:PTR
STATus:OPERation:CONDition?
This query returns the value of the Operation Condition register. That is a read-only register which holds
the real-time (unlatched) operational status of the ac source.
Query Syntax
STATus:OPERation:CONDition?
Parameters
None
Examples
STAT:OPER:COND?
Returned Parameters
<NR1> (register value)
Related Commands
STAT:QUES:COND?
STATus:OPERation:ENABle
This command and its query set and read the value of the Operation Enable register. This register is a
mask for enabling specific bits from the Operation Event register to set the operation summary bit (OPER)
of the Status Byte register. The operation summary bit is the logical OR of all enabled Operation Event
register bits.
Command Syntax
STATus:OPERation:ENABle <NRf+>
Parameters
0 to 32767 | MAXimum | MINimum
Default Value
0
Examples
STAT:OPER:ENAB 32 STAT:OPER:ENAB 1
Query Syntax
STATus:OPERation:ENABle?
Returned Parameters
<NR1> (register value)
Related Commands
STAT:OPER?