Agilent Technologies 6814B Welding System User Manual


 
Language Dictionary - 3
79
LIST:VOLTage:OFFSet:SLEW
Agilent 6811B, 6812B, 6813B, Only
This command specifies the dc offset slew list points. The slew points are given in the command
parameters, which are separated by commas. The order in which the points are entered determines the
sequence in which they are output when a list is triggered. Changing list data while a subsystem is in list
mode generates an implied ABORt.
Command Syntax
[SOURce:]LIST:VOLTage:OFFSet:SLEW <NRf+> | INF {,<NRf+> | INF}
Parameters
0 to 9.9E37 | MAXimum | MINimum | INFinity
Unit
V
(volts per second)
Examples
LIST:VOLT:OFFS:SLEW 10,20,1E2
Query Syntax
[SOURce:]LIST:VOLTage:OFFSet:SLEW?
Returned Parameters
<NR3> {,<NR3>}
Related Commands
LIST:VOLT:SLEW:POIN? LIST:COUN LIST:DWEL LIST:STEP
LIST:VOLTage:OFFSet:SLEW:POINts?
Agilent 6811B, 6812B, 6813B, Only
This query returns the number of points specified in LIST:VOLTage:OFFSet:SLEW. Note that it returns
only the total number of points, not the point values.
Query Syntax
[SOURce:]LIST:VOLTage:OFFSet:SLEW:POINTs?
Returned Parameters
<NR1>
Examples
LIST:VOLT:OFFSet:SLEW:POIN?
Related Commands
LIST:VOLT:OFFS