Agilent Technologies 6814B Welding System User Manual


 
Language Dictionary - 3
97
Bit Configuration of Questionable Status Registers
Bit
Position
15 14 13 12 11 10 9 8–5 4 3 2 1 0
Bit Name not
used
Meas
Ovld
Isum CL
rms
Rail CL
peak
RI not
used
OT UNR SOA OCP OV
Bit
Weight
16384 8192 4096 2048 1024 512 16 8 4 2 1
OV over-voltage protection has tripped
OCP over-current protection has tripped
SOA safe operating area protection has tripped (Agilent 6811B, 6812B, 6813B)
UNR output is unregulated
OT over-temperature protection has tripped
RI remote inhibit is active
CL peak peak current limit is active (Agilent 6811B, 6812B, 6813B)
Rail rail protection tripped (Agilent 6811B, 6812B, 6813B);
rail voltage unregulated (Agilent 6814B, 6834B, 6843A)
CL rms rms current limit is active
Isum summary of Isum registers (Agilent 6834B)
MeasOvld current measurement exceeded low current range capability (Agilent 6811B, 6812B, 6813B)
STATus:QUEStionable?
This query returns the value of the Questionable Event register. The Event register is a read-only register
which holds (latches) all events that are passed by the Questionable NTR and/or PTR filter. Reading the
Questionable Event register clears it.
NOTE: On the Agilent 6834B, each signal that is fed into the Questionable Status Condition
register is logically-ORed from three corresponding status signals that originate from each
phase.
Query Syntax
STATus:QUEStionable[:EVENt]?
Parameters
None
Examples
STAT:QUES:EVEN?
Returned Parameters
<NR1> (register value)
Related Commands
*CLS STAT:QUES:NTR STAT:QUES:PTR
STATus:QUEStionable:CONDition?
This query returns the value of the Questionable Condition register. That is a read-only register which
holds the real-time (unlatched) questionable status of the ac source.
Query Syntax
STATus:QUEStionable:CONDition?
Parameters
None
Examples
STAT:QUES:COND?
Returned Parameters
<NR1> (register value)
Related Commands
STAT:OPER:COND?