CURRent:PEAK
Agilent 6811B, 6812B, 6813B, Only
This command sets the output limit of the absolute value of peak instantaneous current.
Command Syntax
[SOURce:]CURRent:PEAK[:IMMediate] <NRf+>
Parameters
refer to Specifications Table in User’s Guide
Unit
A
(peak amperes)
*RST Value
13 (Agilent 6811B/6812B)
26 (Agilent 6813B)
Examples
CURR:PEAK:IMM 15
Query Syntax
[SOURce:]CURRent:PEAK[:IMMediate]?
Returned Parameters
<NR3>
Related Commands
CURR:PEAK:MODE CURR:PEAK:TRIG
CURRent:PEAK:MODE
Agilent 6811B, 6812B, 6813B, Only
This command determines how the peak current limit is controlled during a triggered output transient. The
choices are:
FIXed
The peak current limit is unaffected by a triggered output transient.
STEP
The peak current limit is programmed to the value set by CURRent:PEAK:TRIGgered
when a triggered transient occurs.
PULSe
The peak current limit is changed to the value set by CURRent:PEAK:TRIGgered for a
duration determined by the pulse commands.
LIST
The peak current limit is controlled by the peak current list when a triggered transient
occurs.
Command Syntax
[SOURce:]CURRent:PEAK:MODE <mode>
Parameters
FIXed | STEP | PULSe | LIST
*RST Value
FIXed
Examples
CURR:PEAK:MODE FIX
Query Syntax
[SOURce:]CURRent:PEAK:MODE?
Returned Parameters
<CRD>
Related Commands
CURR:PEAK CURR:PEAK:TRIG