Agilent Technologies 6814B Welding System User Manual


 
3 - Language Dictionary
78
LIST:VOLTage:SLEW:POINts?
This query returns the number of points specified in LIST:VOLTage:SLEW. Note that it returns only the
total number of points, not the point values.
Query Syntax
[SOURce:]LIST:VOLTage:SLEW:POINTs?
Returned Parameters
<NR1>
Examples
LIST:VOLT:SLEW:POIN?
Related Commands
LIST:VOLT:SLEW
LIST:VOLTageOFFSet
Agilent 6811B, 6812B, 6813B, Only
This command specifies the dc offset points in a list. The offset points are given in the command
parameters, which are separated by commas. The order in which the points are entered determines the
sequence in which they are output when a list is triggered. Changing list data while a subsystem is in list
mode generates an implied ABORt.
The maximum peak voltage that the ac source can output is 425 V peak. This includes any combination of
voltage, voltage offset, and function shape values. Therefore, the maximum value that can be
programmed depends on the peak-to-rms ratio of the selected waveform. For a sinewave, the maximum
voltage that can be programmed is 300 V rms.
Command Syntax
[SOURce:]LIST:VOLTage:OFFSet <NRf+> {,<NRf+>}
Parameters
–425 to +425 | MAXimum | MINimum
Unit
V
(dc voltage)
Examples
LIST:VOLT:OFFS 50,75,100
Query Syntax
[SOURce:]LIST:VOLTage:OFFSet?
Returned Parameters
<NR3> {,<NR3>}
Related Commands
LIST:VOLT:OFFS:POIN? LIST:COUN LIST:DWEL
LIST:STEP LIST:VOLT:SLEW
LIST:VOLTage:OFFSet:POINts?
Agilent 6811B, 6812B, 6813B, Only
This query returns the number of points specified in LIST:VOLTage:OFFSet. Note that it returns only the
total number of points, not the point values.
Query Syntax
[SOURce:]LIST:VOLTage:OFFSet:POINTs?
Returned Parameters
<NR1>
Examples
LIST:VOLT:OFFS:POIN?
Related Commands
LIST:VOLT:OFFS