Teledyne t100h Work Light User Manual


 
13
2. SPECIFICATIONS, APPROVALS & WARRANTY
2.1. Specifications
Table 2-1: Model T100H Basic Unit Specifications
Min/Max Range
(Physical Analog Output)
Min: 0-10 ppm Full scale
Max: 0-5,000 ppm Full scale (selectable, dual ranges and auto ranging supported)
Measurement Units ppm, mg/m
3
(user selectable)
Zero Noise
1
0.1 ppm (RMS)
Span Noise
1
< 1% of reading (RMS) above 10 ppm
Lower Detectable Limit
2
0.2 ppm
Zero Drift (24 hours) < 1 ppm
Span Drift (24 hours) < 0.5% fo full scale
Linearity 1% of full scale
Precision 0.5% of reading above 10 ppm
Voltage Coefficient < 0.05% per V
Lag Time
1
5 seconds
Rise/Fall Time
1
< 30 sec to 95%
Sample Flow Rate 700 cm
3
/min ±10%
AC Power Rating 100V-120V, 220V-240V, 50/60 Hz
Recorder Offset ±10%
Analog Output Ranges mV, 1 V, 5 V, 10V, 5V, 1V, 0.1V (selectable)
Analog Output Resolution 1 part in 4096 of selected full-scale voltage
Included I/O
1 Ethernet: 10/100Base-T
2 RS232 (300-115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs
6 opto-isolated digital control inputs
4 analog outputs
Optional I/O
1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
Temperature Range 5-40
o
C
Temperature Coefficient < 0.1% per
o
C
Humidity Range 0 - 95% RH, non-condensing
Environmental Installation category (over-voltage category) II; Pollution degree 2
Dimensions H x W x D 7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight Analyzer: 31 lbs (16 kg); External pump: 15 lbs (7kg)
Certifications
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class
A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.
IEC 61010-1:90 + A1:92 + A2:95,
1
As defined by the USEPA.
2
Defined as twice the zero noise level by the USEPA.
07265A DCN6038