10.3.2. Sensor Module Repair & Cleaning
NOTE:
After any repair or service has been performed on the sensor module, the T100H should be allowed to
warm up for 60 minutes.
Always perform a perform a leak check (See Section 11.5.1) and calibrate the analyzer (see Section 7)
before placing it back in service.
The most significant difference between the T100 sensor module and the T100H sensor module is the location
of the reference detector. Therefore most of the procedures described in Section 11.6.3 apply to the T100H as
well.
Exceptions are noted below:
Gas Inlet
fitting
TEC Power
Cable
Connector
UV Detector
Housing
Shutter Cabling
Exits here
UV Lamp Power
Suppl
y
Wirin
g
UV Beam
Splitter
Thermistor
Wiring exits here
Heater Wiring
exits here
J5
J6
Gas Outlet
fitting
Figure 10-2: Sensor Module Wiring and Pneumatic Fittings
07265A DCN6038