Teledyne API - T100H Addendum to T100 Operation Manual Getting Started
21
3.3. Initial Operation
With the following exceptions, the operation of the T100H is nearly identical to that of the T100. Please refer to
the T100 User’s Manual, Section 3, for details on initial operation, including common warning messages,
functional check, and initial calibration and common interferents for the T100H.
3.3.1. Warning Messages
Please refer to the T100 User’s Manual (06807), Section 3, for a complete listing of warnings for the T100H.
The following table lists warnings that differ in the T100H from those described in the T100 manual.
Table 2-1: Possible Warning Messages at Start-Up
MESSAGE MEANING
Vacuum Pressure Warning
The vacuum pressure reading is out of its allowed range. The pump may
have failed, or the instrument may have a leak or obstruction in the flow
path.
3.3.2. Test Functions
Check to make sure that the analyzer is functioning within allowable operating parameters as described in the
T100 Manual (P/N 06807). The available test functions for the T100H are:
SAMPLE RANGE = 50.000 PPM SO2=X.XXX
< TST TST > CAL SETUP
Toggle <TST TST> buttons to
scroll through list of functions
RANGE
STABIL
VAC
PRES
SAMP FL
PMT
NORM PMT
UV LAMP
LAMP RATIO
STR. LGT
DARK PMT
DARK LAMP
SLOPE
OFFSET
HVPS
RCELL TEMP
BOX TEMP
PMT TEMP
TEST
1
TIME
1
Only appears if analog output
A3 is actively reporting a test
function
07265A DCN6038