Teledyne t100h Work Light User Manual


 
25
5. OPTIONAL HARDWARE AND SOFTWARE
With the following additions, changes and exceptions, the options listed in Table 1-1 of the T100 Manual (P/N
06807) are also available for the T100H.
5.1. Ambient Zero/Ambient Span Valves (Option 50A)
The T100H zero/span valve option is identical to that of the T100 in respect to operation and valve states (see
Table 3-10 of the T100 Manual). The internal pneumatic connections are slightly different.
FLOW CONTROL &
PRESSURE SENSOR
VACUUM
PRESSURE
SENSOR
SAMPLE
PRESSURE
SENSOR
CRITICAL
FLOW ORIFICE
700 cm
3
/min
@ 7 In-Hg-A
UV
LAMP
PMT
REACTION
CELL
Particulate
Filter
PUMP
EXHAUST
GAS OUTLET
INSTRUMENT CHASSIS
UV
Detector
GAS INPUT
MANIFOLD
Span/Cal
Valve
SAMPLE
GAS
INLET
ZERO AIR
INLET
HIGH
SPAN AIR
INLET
Zero
Gas
Valve
LOW
SPAN AIR
INLET
Low
Span
Valve
High
Span
Valve
Figure 5-1: Pneumatic Diagram of the T100H With Z/S Option Installed.