Teledyne t100h Work Light User Manual


 
Teledyne API - T100H Addendum to T100 Operation Manual Getting Started
17
3.2. Pneumatic Connections
NOTE
To prevent dust from getting into the analyzer, it was shipped with small plugs inserted into each of the
pneumatic fittings on the rear panel. Make sure that all dust plugs are removed before attaching exhaust
and supply gas lines. Store for future use.
Table 3-1: Inlet / Outlet Connector Descriptions
REAR PANEL
LABEL
FUNCTION
Sample
Connects the sample gas to the analyzer. When operating the analyzer without zero
span option, this is also the inlet for any calibration gases.
Exhaust Connect an exhaust gas line to this port to the inlet of the external pump.
Zero Air
On Units with zero/span valve option installed, this port connects the zero air gas or the
zero air cartridge to the analyzer.
Figure 3-5 of the T100 Manual (P/N 06807) shows the internal pneumatic flow of the T100 in its standard
configuration. For a diagram of the internal pneumatic flow of the T100H, see Figure 3-2 of this addendum.
3.2.1.1. Pneumatic Connections to T100H Basic Configuration:
The pneumatic connections for the T100H analyzer in its basic configuration are nearly identical to those
described the T100 Manual (P/N 06807) Section 3.5 except that the T100H has an external pump. Therefore:
A pneumatic line of ¼” PTEF must be attached between the analyzer’s exhaust port and the inlet port of
the pump.
The exhaust from must be vented outside the shelter or immediate area surrounding the instrument
using a maximum of 10 meters of 1/4” PTFE tubing.
Source of
SAMPLE Gas
Removed
during
Calibration
MODEL 701
Zero Air
Generator
MODEL T700
Gas Dilution
Calibrator
PUMP
vented outside immediate area
(VENT
if not
venting at
calibrator)
VENT here
if sample gas is
supplied under
pressure
Chassis
Sample
EXHAUST
Span
Zero Air
Calibrated SO
2
at HIGH
Concentration
Figure 3-1: Example of Pneumatic Connections to T100H External Pump
This change is true for all configurations and variations of the T100H.
07265A DCN6038