TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual)
52
20. Adjust gain adjustment potentiometer until the NORM PMT value is ±10 mV of the target level from Step 16.
21. Perform span calibration (see Section 7 of the T100 Manual)
22. Scroll to the SLOPE function and record the value.
23. If the value of the SLOPE is between 0.900 and 1.100 the PMT Hardware calibration is complete.
24. If the value of the SLOPE is less than 0.900 or greater than 1.100:
1. Multiply the slope value from step 22 by the norm PMT value from step 19.
2. Repeat steps 17 through 24 using this new value for NORM PMT.