Appendix C
Warranty/Repair Questionnaire
Models T100H, 100EH
(05029D DCN 6038)
TELEDYNE API CUSTOMER SERVICE
EMAIL: api-customerservice@teledyne.com
PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816
C-1
CUSTOMER: ____________________________________ PHONE: ___________________________________________
CONTACT NAME: _______________________________ FAX NO. ___________________________________________
SITE ADDRESS: ______________________________________________________________________________________
MODEL 100EH SERIAL NO.: ________________________FIRMWARE REVISION: _______________________________
1. ARE THERE ANY FAILURE MESSAGES? ____________________________________________________________
_____________________________________________________________________________________________________
_____________________________________________________________________________________________________
PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST
PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT)
Parameter Displayed As Observed Value Units Nominal Range
Range
RANGE
PPM
UG/M
3
1-5000 PPM Standard
Stability
STABIL
PPM
UG/M
3
<.1 PPM with Zero Air
Vacuum
VAC
In-Hg-A
4 – 10 “Hg
Sample Pressure
PRES
In-Hg-A 24 – 29
Sample Flow
SAMP FL
CC/MIN
700 ±10%
PMT Signal
PMT
MV
0 ± 100 with Zero Air
Normalized PMT Signal
NORM PMT
MV
0 ± 100 with Zero Air
UV Lamp
UV LAMP
MV 1000 – 4800
UV Lamp Ratio
LAMP
RATIO
% 35 – 120%
Stray Light
STR. LGT
PPM -50 to +100
Dark PMT
DRK PMT
MV <200
Dark Lamp
DRK LMP
MV -30 to 50
Slope
SLOPE
- 1.0 ± 0.3
Offset
OFFSET
MV <200
High Voltage Power Supply
HVPS
V 400 – 750*
Reaction Cell Temperature
RCELL
TEMP
°C 50 ± 1
Box Temperature
BOX TEMP
°C
Ambient + (3-7)
PMT Temperature
PMT TEMP
°C 7 ± 2
Time of Day
TIME
HH:MM:SS
07265A DCN6038