104 Hearing Aid Measurements WinCHAP User’s Manual
7.5.2 Starting a test sequence
If your analyzer is already set up for testing, then you can remotely start the
automated test sequence from WinCHAP by clicking Start Test
:
1. Set the analyzer up for normal coupler testing.
2. Click Start Test in the automated test sequence in WinCHAP. This will
remotely start the automated test sequence on your analyzer.
3. If the analyzer pauses midway through the test to allow for adjustment of
the hearing aid to reference test gain (or to put the aid in telecoil mode),
a window will pop up on WinCHAP, telling you what to do. Adjust the
hearing aid accordingly.
4. Click
OK in the test window when you have done the adjustment.
WinCHAP will tell the analyzer to resume testing. When the test sequence
is done, WinCHAP will display the results.
5. Click
Save in the toolbar to save the results.