4 Move down to the “parameters” level.
^000000:CODE
0 3 3
1 2 0 5 Unsecure the function generator by entering the security code.
ENTER
^033120:CODE
The security code is set to “HP33120” when the function generator is
shipped from the factory. The security code is stored in non-volatile
memory and does not change when the power has been off or after a
remote interface reset.
To enter the security code from the front panel, enter only the six digits.
To enter the security code from the remote interface, you may enter up to
12 characters. Use the knob or arrow keys to move left or right between
digits. Use the up or down arrow keys to change the digits.
To re-secure the function generator following a calibration, perform this
procedure again.
Additional information about the calibration security feature is given
on page 64.
Chapter 3 Front-Panel Menu Operation
To unsecure the function generator for calibration