General Calibration/Adjustment Procedure
The adjustment procedures described in chapter 4 use the CAL MENU to
generate and set internal calibration constants. The general menu
procedure is the same for all calibration setups. The following example
demonstrates making the Frequency and Burst Rate adjustments.
Shift 1 Turn on the menu.
Menu On/Off A: MOD MENU
< 2 Move across to the CAL MENU choice on this level.
1
F: CAL MENU
3 Move down a level to the UNSECURED command.
1: UNSECURED
If the display shows SECURED, you will have to unsecure the function
generator to perform the calibration procedures. A procedure is given on
page 47.
> 4 Move across to the CALIBRATE choice.
1
2: CALIBRATE
5 Move down one level.
The display indicates the calibration setup number. You can change this
number to perform individual specification adjustments.
SETUP 00
Enter 6 Begin the Frequency and Burst Rate adjustment procedure.
4
Chapter 4 Calibration Procedures
General Calibration/Adjustment Procedure
67