< > 7 Move the flashing cursor over the digit to be edited.
1
^
8 Change the value in the display to match the measured frequency.
1.000,0040KHz
Enter 9 Calculate and save the new value.
CALIBRATING
10 Perform the next adjustment procedure.
The setup number and function generator output is automatically set for
the next adjustment procedure.
SETUP ^01
You will press ENTER twice for each calibration step, once to select
the setup (as described in step 6) and once to enter the adjustment
(as described in step 9).
1
You can also use the knob to scroll left or right through the choices on each level of
the menu
Chapter 4 Calibration Procedures
General Calibration/Adjustment Procedure
68