Agilent Technologies 33120A Welding System User Manual


 
4 Use the DMM to measure the function generator ACrms output voltage
for each setup in the table on the next page. These adjustments use the
50
W load and cable measured in step 2 and connected as shown below.
Chapter 4 Calibration Procedures
AC Amplitude Adjustment (50
W)
74