Install the 50W feedthrough load between the DMM and the function
generator output. Set the function generator for each output indicated in
the table on the next page. Use a DMM to measure the ACrms output
voltage of the function generator. Compare the measured results to the
test limits shown in the table. This is a 50
W output termination test.
Agilent 33120A Measurement
Function OUT TERM
1
Ampl Freq Nominal Error
Q Sine wave
50 W
3.5 Vrms 1.0000 kHz 3.5 Vrms
0.035 Vrms
Sine wave
50 W
2.8 Vrms 1.0000 kHz 2.8 Vrms
0.028 Vrms
Sine wave
50 W
2.2 Vrms 1.0000 kHz 2.2 Vrms
0.022 Vrms
Sine wave
50 W
1.7 Vrms 1.0000 kHz 1.7 Vrms
0.017 Vrms
Sine wave
50 W
1.4Vrms 1.0000 kHz 1.4Vrms
0.014 Vrms
Sine wave
50 W
1.1 Vrms 1.0000 kHz 1.1 Vrms
0.011 Vrms
Q Sine wave
50 W
0.88 Vrms 1.0000 kHz 0.88 Vrms
0.0088 Vrms
Sine wave
50 W
0.70 Vrms 1.0000 kHz 0.70 Vrms
0.0070 Vrms
Sine wave
50 W
0.55 Vrms 1.0000 kHz 0.55 Vrms
0.0055 Vrms
Sine wave
50 W
0.44 Vrms 1.0000 kHz 0.44 Vrms
0.0044 Vrms
Sine wave
50 W
0.35 Vrms 1.0000 kHz 0.35 Vrms
0.0035 Vrms
Sine wave
50 W
0.28 Vrms 1.0000 kHz 0.28 Vrms
0.0028 Vrms
Sine wave
50 W
0.22 Vrms 1.0000 kHz 0.22 Vrms
0.0022 Vrms
Sine wave
50 W
0.17 Vrms 1.0000 kHz 0.17 Vrms
0.0017 Vrms
Sine wave
50 W
0.14 Vrms 1.0000 kHz 0.14 Vrms
0.0014 Vrms
Sine wave
50 W
0.11 Vrms 1.0000 kHz 0.11 Vrms
0.0011 Vrms
Sine wave
50 W
0.088Vrms 1.0000 kHz 0.088Vrms
0.00088 Vrms
Sine wave
50 W
0.070 Vrms 1.0000 kHz 0.070 Vrms
0.00070 Vrms
Sine wave
50 W
0.055 Vrms 1.0000 kHz 0.055 Vrms
0.00055 Vrms
Sine wave
50 W
0.044 Vrms 1.0000 kHz 0.044 Vrms
0.00044 Vrms
Q Sine wave
50 W
0.035 Vrms 1.0000 kHz 0.035 Vrms
0.00035 Vrms
Sine wave
50 W
0.028 Vrms 1.0000 kHz 0.028 Vrms
0.00028 Vrms
Sine wave
50 W
0.022 Vrms 1.0000 kHz 0.022 Vrms
0.00022 Vrms
Sine wave
50 W
0.018 Vrms 1.0000 kHz 0.018 Vrms
0.00018 Vrms
1
Output termination set using front panel controls. HIGH Z assumes no load on
output. 50W assumes a 50W 0.1W load on output.
4
Chapter 4 Calibration Procedures
AC Amplitude Verification