Table 1-2. HP 16532A Operating Characteristics (cont.)
WAVEFORM DISPLAY
Displayed Waveforms:
Eight waveform windows maximum, with scrolling across 96 waveforms.
Display Formats:
Waveforms can be displayed in an overlapping and/or non-overlapping format.
Display capability of A B and A+ B is also provided.
Display Resolution:
500 points horizontal, 240 points vertical
DISPLAY MODES
Normal (Single):
New acquisition replaces old acquisition on screen.
Accumulate:
New acquisition is displayed in addition to previous acquisitions until screen is
erased.
Average:
New acquisitions are averaged with old acquisitions, with the updated waveform
displayed until erased. Maximum number of averages is 256.
Overlay:
Up to 8 acquired waveforms from separate channels can be overlayed in the
same display area.
Connect-the-dots:
Provides a display of the sample points connected by straight lines.
Waveform Reconstruction:
When there is insufficient data to fill every horizontal location, a post acquisition
reconstruction filter fills in missing data points for time base < 50 ns/div.
Waveform Math:
Display capability of A B and A+ B functions is provided.
HP 16532A - General Information