5. On the HP 54502A Oscilloscope, set the following parameters, in the order given.
Menu Selection Setting
Channel
Timebase
Trigger
Probe
V/Div
Offset
Coupling
s/Div
Acquisition
Mode
Level
Slope
1.000:1
200 mV
0.0 V
DC
10 ns
Realtime
Edge
− 200 mV
Falling
6. In the HP 16532A Oscilloscope trigger menu, select Auto-Trig On. Verify that the 100 MHz System
Clock is gated.
7. Select STOP on both the HP 54502A and HP 16532A oscilloscopes.
8. Disconnect the cable from both oscilloscopes.
HP 16532A - Service